Open MRI System
Oper
Technical Specifications of OPER-0.4 MRI System
Magnet
- Type: Permanent magnet, eddy current free, open C type & self-temperature control
- Field strength: 0.4T±3%
- Field homogeneity:
- <±2ppm(FWHH)/20cm DSV
- <5ppm(VRMS)/40cm DSV
- Vertical gap distance: 40cm
Gradient system
Amplifier:
- Field strength: single axis 25mT/m
- Gradient rise time:0.33ms
- Slew rate: single axis , 75.7mT/m/ms
- Nonlinearity: 40cm*40cm*38cm ellipsoid<5%
RF transmitting & receiving system
- Type of transmitter & receiver: Digital
- Frequency range: DC-22MHZ
- Maximum receiver bandwidth: 1.25MHZ
- Pre-amplifier: gain 30dB
- Noise figure: 0.3dB
- RF amplifier: ≥6kw
- RF transmitting coil: panel transmitting coil
- RF receiving coil: quadrature coil
Auto pre-scan function
Auto RF correction, auto frequency lock, auto gain, auto coil identification, auto coil tuning, active shim
Image specifications
- Image types: T1 weighted image, T2 weighted image, T2* weighted image, PD weighted image, diffusion weighted image, fat suppressed image, water suppressed image, Magnetic resonance hygrography (MRCP MRM MRU), MRA
- Field of view: 40mm-400mm
- Acquisition matrix: 64x64-512x512
- Graphics matrix max.: 1024x1024
- Body in 30cm FOV: 1.5mm
- Head in 24cm FOV: 1mm
- Slice orientation: transverse, sagittal, coronal, free oblique, multi-slice & multi-range
- Slice gap: 0-20mm
Pulse sequence
- Spin echo (SE)
- Multi-slice & Multi-echo (MSME)
- Fast Spin Echo (FSE)
- Single shot fast spin echo (SSFSE)
- Multi shot fast spin echo (MSFSE)
- Inversion recovery (IR)
- Short inversion tau IR (STIR)
- Fluid attenuated IR (FLAIR): T1W, T2W
- Inversion recovery fast spin echo (IRFSE)
- Gradient echo (GRE): 2D/3D
- Steady state process GRE (SSPGRE)
- MRA: TOF 2D/3D
- Echo Planar Imaging (EPI)
Image display & post processing
- Image post processing: filter, zoom, cut, text illustration & elimination, image offset, auto adjustment of window length & width
- Image analysis: interested area illustration, distance measurement
Patient Management
- Talkback system
- Patient bed bearing - 240Kg Max.
Computer system
- Main: dual-core processor (≥2.8GHz)
- Network controller: 1000Mbps
- Memory: 2 GB
- Graphics memory: 128MB
- Hard driver: 160 GB
- Removable storage: DVD+RW
- Image reconstruction: (256x256) 285 Frames/s
- 21” TFT monitor with at least 1280x1024 resolution
DICOM 3.0 with compatibility